By contrast, the multifocal VEP (MVEP) is a rapidly evolving technique for assessing visual function and detecting visual field defects.4-13 This technology has been adapted for use in the AccuMap (ObjectiVision Pty Ltd, Sydney, Australia; distributed in the US by Heidelberg Engineering GmbH, Dossenheim, Germany) to provide an objective means of assessing the visual field (Figure 1).14,15

Figure 1. The AccuMap V2.0 Objective Perimeter.
MVEP TECHNIQUE
Baseler et al16 first reported the recording of an MVEP. They used the VERIS system (Electro-Diagnostic Imaging, San Mateo, CA), which employs pseudo-randomly presented multifocal stimulation together with cortical scaling of the size of the stimulated patches. The subject viewed a screen displaying a rapidly alternating checkerboard pattern with multiple zones, each changing according to a different sequence in time. Single-channel recordings were made via scalp electrodes. Individual VEPs from each zone were extracted from the noise in the form of binary kernels by means of applied mathematics, specifically, a Fast Walsh Transform. The investigators concluded that the MVEP could not be applied clinically due to the high variability of responses between individuals.
We modified the technique.7 We positioned a bipolar electrode 2 cm above and 2 cm below the inion (the bony protuberance on the occipital skull) and recorded MVEPs from 60 areas of the visual field up to 26º of eccentricity in normal subjects. The electrode configuration was then further enhanced to a four-channel system employing an occipital cross-electrode holder. The plastic cross provided a curved, flexible support for four gold-cup electrodes attached to a headband. When centered over the inion, the recording positions were 2 cm above, 6 cm below, and 4 cm to either side. The electrodes could be linked to provide vertical, horizontal, and two oblique bipolar recording channels. This setup provided better signals from many zones (particularly along the horizontal meridian) that were missed if only a vertically oriented channel were used. The cross helped reduce variability between subjects and standardized the electrode positions to improve reproducibility.9 The stimulus check size and cortical scaling were confirmed as optimal for signal detection across the field.17
Because the test's application was still limited by interindividual variability, a means of compensation was required. During the recording, we applied a scaling algorithm based on underlying electroencephalogram (EEG) amplitudes, which reflect the characteristic signal conduction for each individual.18 This approach was based on the assumption that the principal factors governing the variability of signal amplitude between individuals were (1) the conductivity of the skull, subcutaneous tissues, and scalp and (2) such conductivity's proportional effect on the EEG, which is a much larger signal. Once alpha rhythm spikes (further information presented later) and electrocardiogram contamination are identified and removed on Fourier analysis, a scaling factor is applied to the MVEP data based on normative data. With this technique, interindividual variability dramatically decreased, thus enabling the creation and use of a normals database for the identification of localized field loss.
An alternative method employed by Hood et al19 involved examining the signal-to-noise ratio (SNR) rather than amplitudes. Although a reasonable option, it is more affected by background alpha rhythm, because the frequency of alpha rhythm lies in the 8- to 12-Hz range, where the majority of the MVEP signal occurs. Alpha rhythm increases when a subject loses visual concentration, but this finding can be normal in some individuals, even when they are alert. We have found that the use of an interactive fixation target (such as is used in the AccuMap) reduces alpha rhythm levels in normal subjects.20 The other important factor is that, with the EEG scaling method, examiners can identify the alpha rhythm and other contaminants such as electrocardiogram signal (ECG) on the Fourier spectrum; they can then remove contaminants prior to determining the scaling factor to be applied for each individual. This ability provides a distinct advantage over the SNR technique, because the presence of contaminants reduces the final SNR determined, without any compensation for the overall size of the signal.
The AccuMap provides a Fourier analysis of frequencies after each run, and it allows operators to view compressed data so that they can exclude noisy runs. Cumulative runs are added until the signal is stable, and the software alerts the technician when sufficient runs have been collected. In contrast, when using VERIS, the entire 7- or 15-minute run must be completed before any data are available for analysis. The AccuMap provides alerts to the technician for high- or low-frequency noise and alpha rhythm. The new V2.1 of the device will provide a noise index as a guide.
SENSITIVITY IN GLAUCOMA
In a study using the AccuMap V1.3, Goldberg et al14 reported greater than 95% sensitivity in glaucoma patients (sample of 100 eyes of 100 subjects, the visual field mean deviation for the glaucoma group was -6.5 dB). Many fellow eyes with normal visual fields had MVEP defects. In a further study of 436 consecutive patients, investigators reported similar sensitivity of 92% in early glaucoma and of 100% in moderate glaucoma (mean deviation > 6 dB).21 More importantly, the MVEP was extremely helpful in assessing patients with unreliable or variable subjective fields and for cases in which the degree of field loss was out of proportion to the disc changes. Using AccuMap V1.3, Fortune et al15 found that, with a specificity of 90%, the test's sensitivity was between 78% and 97%, depending on the criteria used for abnormality. Similar sensitivity levels have been reported using a multichannel MVEP on the VERIS system to look at unilateral hemifield defects.22
Figure 2 shows an example of an MVEP recording and the corresponding subjective visual field result. The area of the scotoma corresponds well between the subjective Humphrey visual field (Carl Zeiss Meditec Inc., Dublin, CA) and the AccuMap probability plot. Figure 3 shows a different patient who underwent the test five times on 5 different days, each 1 week apart, and whose scotomas were consistently identified on the AccuMap. Between-test variability of the MVEP signal on the AccuMap V1.3 is approximately 15%,14 but recently conducted studies using the AccuMap V2.0 in 50 subjects tested twice demonstrated a mean point variability of 12% (Graham and Klistorner, unpublished data). Chen et al23 found that the test/retest reliability of the MVEP was superior to that of subjective perimetry when a logarithmic scale was used.

Figure 2. This MVEP recorded with the AccuMap system shows a superior field defect in the eye of a patient with glaucoma. VEP signals from all 58 zones (central rings expanded for presentation) is shown (top, left). An amplitude deviation probability plot based on significant differences from the normals database (100 subjects) is shown (top, right). The lower figures show the corresponding Humphrey visual field grey scale and pattern deviation plot.

Figure 3. This example demonstrates the reproducibility of multifocal VEP recording in a glaucoma patient with a superior field defect on Humphrey testing (upper row). The AccuMap identifies the defect but also shows an early inferior arcuate defect. This patient underwent testing five times on different days 1 week apart. Despite some variability in the probability levels of the scotoma between visits, the distribution of field loss is consistent.
Comparing VEP waveforms and the amplitude of the signals between the two eyes of the same individual is another useful technique for detecting early abnormality.6,10 Although different parts of the retina are being stimulated in the two eyes, the information from similar parts of the visual field of both eyes projects to identical areas of the visual cortex, and the resulting VEP waveforms are virtually identical. This technique has limited value in more advanced disease, because scotomas may appear in the same area of the visual field in the two eyes (eg, bilateral superior arcuates). In these cases, it is necessary to rely on the monocular deviation plot.
CORRELATION WITH STRUCTURAL CHANGES
Objective perimetry has demonstrated a good correlation with structural changes as detected by the scanning laser ophthalmoscope (HRT2; Heidelberg Engineering GmbH), both in patients with glaucoma (n=125) and in glaucoma suspects (n=126).24 The agreement between the tests was 81%. Using the AccuMap's severity index and the Moorfields classification on the HRT, all glaucoma subjects were identified by either one or both types of objective tests. MVEP signals are affected by media opacity in the central two rings of the test zones, because the check sizes are very small in these areas. A study by Whitehouse25 of 15 cataract patients tested pre- and postoperatively demonstrated an improvement in the signal from the central rings, but the peripheral zones did not change significantly.
The MVEP can also be employed in neuro-ophthalmic disorders such as optic neuropathy, cortical lesions, and functional loss.26-29 Additionally, it is also useful in the assessment of children, in whom it is best restricted to short sequences of stimulus presentation.30
CONCLUSION
The MVEP is now a well-characterized electrophysiologic response to demonstrate localized visual field loss in the clinical setting. The technique is not specific to ganglion cells, because it reflects damage anywhere along the visual pathway. It can, however, be adapted with multichannel recording and appropriate EEG-based scaling to identify focal losses and to reliably detect glaucoma. The objective nature of this type of test has distinct clinical advantages over subjective methods of assessing glaucoma; of particular value are the facts that it does not involve a learning curve for the patient and it eliminates indecision on the part of the test subject. Patients also report that they find objective perimetry much less stressful to perform than white-on-white or blue-yellow perimetry. Its noninvasive nature has led to excellent acceptance among our patients. Technicians do require training but do not require a background in electrophysiology in order to perform the test. The scalp electrodes need cleaning between uses, as is the case for standard VEP recording.
Full test time with the AccuMap, including set-up, is approximately 30 minutes per patient. The device can provide an objective assessment of the patient's visual field, and it is particularly useful in subjects who have difficulty with conventional field testing. Studies currently being initiated will determine the technology's role in early detection in preperimetric cases.
Stuart L. Graham, MBBS, MS, FRANZCO, is Clinical Lecturer at the Save Sight Institute at Sydney University in Australia. He disclosed that he is a consultant for ObjectiVision Pty Ltd and a coauthor of two patents licensed to that company. Dr. Graham may be reached at +61 29 231 1833; stuart@eye.usyd.edu.au.
Alex Klistorner, MD, PhD, is Research Officer at the Save Sight Institute at Sydney University in Australia. He disclosed that he is a consultant for ObjectiVision Pty Ltd and a coauthor of two patents licensed to that company. Dr. Klistorner may be reached at +61 29 382 7305; sasha@eye.usyd.edu.au.
1. Bach M. Electrophysiological approaches for early detection of glaucoma. Eur J Ophthalmol. 2001;11(suppl 2):41-49.
2. Fortune B, Johnson CA, Cioffi GA. The topographic relationship between multifocal electroretinographic and behavioural perimetric measures of function in glaucoma. Optom Vis Sci. 2001;78:206-214.
3. Klistorner A, Graham SL, Martins A. Multifocal pattern electroretinogram does not demonstrate localised field defects in glaucoma. Doc Ophthalmol. 2000;100:155-165.
4. Graham SL, Klistorner A. The diagnostic significance of the multifocal pattern visual evoked potential in glaucoma. Curr Opin Ophthalmol. 1999;10:140-146.
5. Graham SL, Klistorner A, Grigg JR, et al. Objective perimetry in glaucoma-recent advances using multifocal stimuli. Surv Ophthalmol. 1999;43(suppl 1):199-209.
6. Graham SL, Klistorner A, Grigg JR, et al. Objective VEP perimetry in glaucoma - Asymmetry analysis to identify early deficits. J Glaucoma. 2000;9:10-19.
7. Klistorner AI, Graham SL, Grigg JR, et al. Multifocal topographic visual evoked potential: improving objective detection of local visual field defects. Invest Ophthalmol Vis Sci. 1998;39:937-950.
8. Klistorner AI, Graham SL, Grigg JR, et al. Electrode position and the multi-focal visual-evoked potential: Role in objective visual field assessment. Aust N Z J Ophthalmol. 1998;26:91-94.
9. Klistorner AI, Graham SL. Objective perimetry in glaucoma. Ophthalmology. 2000;107:2283-2299.
10. Hood DC, Zhang X, Greenstein, VC et al. An interocular comparison of the multifocal VEP: a possible technique for detecting local damage to the optic nerve. Invest Ophthalmol Vis Sci. 2000;41:1580-1587.
11. Hood DC. Assessing retinal function with the multifocal technique. Prog Retin Eye Res. 2000;19:607-646.
12. Hasegawa S, Abe H. Mapping of glaucomatous visual field defects by multifocal VEPs. Invest Ophthalmol Vis Sci. 2001;42:3341-3348.
13. Betsuin Y, Mashima Y, Ohde H, et al. Clinical application of the multifocal VEPs. Curr Eye Res. 1999;22:54-63.
14. Goldberg I, Graham SL, Klistorner A. Multifocal objective perimetry in the detection of glaucomatous field loss. Am J Ophthalmol. 2002;133:29-39.
15. Fortune B, Goh K, Demirel S, et al. Detection of glaucomatous field loss using Multifocal VEP. In: IPS Proceedings 2002/2003. Kugler Publications: The Hague; 2003: 251-260.
16. Baseler HA, Sutter EE, Klein SA, et al. The topography of visual evoked response properties across the visual field. Electroencephalogr Clin Neurophysiol. 1994;90:65-81.
17. Balachandran C, Klistorner A, Graham SL. Effect of stimulus check size on multifocal visual evoked potentials. Doc Ophthalmol. 2003;106:183-188.
18. Klistorner A, Graham SL. Electroencephalogram-based scaling of multifocal visual evoked potentials: Effect on intersubject amplitude variability. Invest Ophthalmol Vis Sci. 2001;42:2145-2152.
19. Zhang X., Hood DC, Chen CS, et al. A signal-to-noise analysis of multifocal VEP responses: an objective definition of poor records. Doc Ophthalmol. 2002;104:303-320.
20. Martins A. Effect of fixation tasks on multifocal VEPs. Paper presented at: The ARVO Annual Meeting; April 29, 2004; Fort Lauderdale, FL.
21. Graham SL, Klistorner A, Goldberg I. Application of multifocal VEP objective perimetry to glaucoma management-clinical experience. Paper presented at: The ARVO Annual Meeting; April 26, 2004; Fort Lauderdale, FL.
22. Thienprasidhi P, Greenstein VC, Chen C, et al. Multifocal visual evoked potential responses in glaucoma patients with unilateral hemifield defect. Am J Ophthalmol. 2003;136:34-40.
23. Chen CS, Hood DC, Zhang X, et al. Repeat reliability of multifocal visual evoked potentials in normal and glaucomatous eyes. J Glaucoma. 2003;12:399-408.
24. Graham SL, Klistorner A, Balachandran C, et al. Comparison of objective tests in glaucoma – Multifocal Objective Perimetry and Heidelberg Retinal Tomography. Paper presented at: The ARVO Annual Meeting; May 7, 2002; Fort Lauderdale, FL.
25. Whitehouse GM. The effect of cataract on Accumap multifocal objective perimetry. Am J Ophthalmol. 2003;136:209-212.
26. Hood DC, Odel JG, Zhang X. Tracking the recovery of local optic nerve function after optic neuritis: a multifocal VEP study. Invest Ophthalmol Vis Sci. 2000;41:4032-4038.
27. Miele DL, Odel JG, Behrens MM, et al. Functional bitemporal quadrantopia and the multifocal visual evoked potential. J Neuro-Ophthalmol. 2000;20:159-162.
28. Odel JG, Winn BJ, Chen CS, et al. The multifocal visual evoked potential in a neuro-ophthalmology practice. Paper presented at: The ARVO Annual Meeting; May 6, 2003; Fort Lauderdale, FL.
29. Wall M, Woodward K, Sleep T. Multifocal visual evoked potential in optic neuropathies and homonymous hemianopias. In: IPS Proceedings 2002/2003. Kugler Publications: The Hague; 2003: 265-274.
30. Balachandran C, Klistorner A, Billson FA. Multifocal VEP perimetry in children: its maturation and clinical application. Br J Ophthalmol. 2004;88:226-232.