OCT using 1.3-µm–wavelength light is better suited for imaging the anterior chamber angle due to two significant properties. First, the amount of scattering in tissue is lower at this wavelength, a difference that increases the penetration of light through scattering ocular structures such as the sclera and the iris so that the morphology of the anterior chamber angle is visualized with more detail (Figure 1). Second, because water in the ocular media strongly absorbs 1.3-µm–wavelength light, only 10% of the light incident on the cornea reaches the retina. This improved retinal protection allows the use of higher-powered illumination, which in turn enables high-speed imaging. The advantages of high-speed imaging include an elimination of motion artifacts, decreased examination time, and an ability to image dynamic ocular events. This article explores how anterior segment OCT works and its potential applications.

Figure 1. The difference in visualization with 0.8-µm– and 1.3-µm–wavelength anterior segment OCT is evident.
Anterior segment OCT has several advantages over gonioscopy for the evaluation of the anterior chamber angle. The latter is subjective, requires a highly trained examiner, and involves ocular contact. In contrast, anterior segment OCT quantitatively measures the anterior chamber angle in a noncontact fashion, and the technology is easy to use after minimal training. Because it does not require contact with the eye, anterior segment OCT is safer and more comfortable for patients, and it avoids the mechanical distortion of the structures to be examined. A potentially significant advantage of this technology is that it may be used in complete darkness. Any changes in the angle configuration that are induced by ambient lighting are thus eliminated.
COMPARING MODALITIES
OCT has several advantages over other cross-sectional imaging modalities such as ultrasound biomicroscopy (UBM) and Scheimpflug photography. Compared with UBM, OCT provides a higher resolution, is completely noncontact, and is relatively easy to perform. Because Scheimpflug photography does not image the actual angle recess, it does not provide visualization of important structural detail. In contrast, OCT provides detailed direct imaging of the anterior chamber angle region.
PROTOTYPES
Several anterior segment OCT prototypes have been described in the literature,4-8 and an FDA-approved device (Visante; Carl Zeiss Meditec Inc., Dublin, CA) recently became commercially available. We have employed some of these prototypes in our studies. The system that we currently use was developed by Carl Zeiss Meditec Inc. It uses a 1.3-µm–wavelength light source and delivers 2,000 A-scans per second. The device acquires and displays eight image frames per second in real time, each with 500 axial scans. The geometry of the scan is telecentric with adjustable scanning widths of 1 to 16 mm and scanning depths of 1 to 8 mm. The axial resolution is 15 µm.
QUANTITATIVELY MEASURING THE ANTERIOR CHAMBER ANGLE
Structures in the anterior chamber angle are well delineated by 1.3-µm OCT (Figure 2). To perform quantitative measurements, the OCT image must first be processed9 to obtain correctly dimensioned images, with adjustments for the geometry of the scan and the refraction of the OCT beam at the eye's anterior surface. Highly reflective and easily identifiable in OCT images, the scleral spur is the landmark used for measuring the parameters of the anterior chamber angle. The quantitative parameters that we have used in our studies include the angle-opening distance at 500 µm, the angle recess area at 500 and 750 µm, and the trabeculo-iris–space area at 500 and 750 µm (Figure 3).

Figure 2. This OCT image of the anterior segment illustrates the anterior chamber angle anatomy. The scleral spur, ciliary body band, angle recess, and iris root are clearly identifiable. (Reprinted with permission from Radhakrishnan S, Huang D, Smith SD. Optical coherence tomography imaging of the anterior chamber angle. Ophthalmol Clin North Am. 2005;18:375-381.)

Figure 3. This anterior segment OCT image demonstrates the quantitative parameters of the anterior chamber angle. AOD represents the angle-opening distance, ARA signifies the angle recess area, and TISA stands for the trabeculo-iris–space area. (Reprinted with permission from Radhakrishnan S, Huang D, Smith SD. Optical coherence tomography imaging of the anterior chamber angle. Ophthalmol Clin North Am. 2005;18:375-381.)
CLINICAL STUDIES
In two recently published studies,8,10 the quantitative measurement of the anterior chamber angle with OCT correlated well with gonioscopy in terms of identifying occludable angles. In one of these studies,8 the investigators also acquired UBM measurements, and they found that both the OCT and UBM parameters were comparable in terms of their reproducibility and correlation with gonioscopy. The best OCT parameters were slightly superior to UBM, with 100% sensitivity and 95.7% specificity for detecting gonioscopically occludable angles.
Anterior segment OCT can quantify changes in the width of the anterior chamber angle due to various factors. A recent case report used a prototypic system to demonstrate the deepening of the anterior chamber angle after a laser peripheral iridotomy.11 Preliminary data from an ongoing study to evaluate illumination-induced changes in the anterior chamber angle with OCT indicate a significant deepening of all anterior chamber parameters under conditions of bright illumination.12
CONCLUSION
Anterior segment OCT appears to be a promising tool for evaluating the anterior chamber angle configuration, including changes induced by illumination and laser peripheral iridotomy. As a rapid, easy-to-use, and completely noncontact modality, the device may be a useful screening tool for occludable angles. Ongoing clinical studies will help to evaluate its efficacy in this regard. The use of anterior segment OCT for evaluating trabeculectomy blebs is a future area of study.
Sunita Radhakrishnan, MD, is Resident, PGY-4, at the Cole Eye Institute, Cleveland Clinic. She has received research support from and served as a consultant to Carl Zeiss Meditec Inc. Dr. Radhakrishnan may be reached at (216) 444-4821; radhaks@ccf.org.
Scott D. Smith, MD, MPH, is Associate Staff, Glaucoma Division, Cole Eye Institute, Cleveland Clinic. He has received research support from Carl Zeiss Meditec Inc. Dr. Smith may be reached at (216) 444-4821; smiths@ccf.org.
1. Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science 1991;254:1178-1181.
2. Swanson EA, Izatt JA, Hee MR, et al. In vivo retinal imaging by optical coherence tomography. Opt Lett. 1993;21:1864-1866.
3. Hee MR, Izatt JA, Swanson EA, et al. Optical coherence tomography of the human retina. Arch Ophthalmol. 1995;113:325-332.
4. Izatt JA, Hee MR, Swanson EA, et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994;112:1584-1589.
5. Koop N, Brinkman R, Lankenau E, et al. Optical coherence tomography of the cornea and the anterior eye segment. Ophthalmologe. 1997;94:481-486.
6. Hoerauf H, Wirbelauer C, Scholz C, et al. Slit-lamp-adapted optical coherence tomography of the anterior segment. Graefe's Arch Clin Exp Ophthalmol. 2000;238:8-18.
7. Radhakrishnan S, Rollins AM, Roth JE, et al. Real-time optical coherence tomography of the anterior segment at 1310nm. Arch Ophthalmol. 2001;119:1179-1185.
8. Radhakrishnan S, Goldsmith J, Westphal V, et al. Comparison of coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;128:1053-1059.
9. Westphal V, Rollins AM, Radhakrishnan S, Izatt JA. Correction of geometric and refractive image distortions in optical coherence tomography applying Fermat's principle. Opt Express. 2002;10:397-404.
10. Wirbelauer C, Karandish A, Haberle H, Pham DT. Noncontact goniometry with optical coherence tomography. Arch Ophthalmol. 2005;123:179-185.
11. Chalita MR, Li Y, Smith S, et al. High-speed optical coherence tomography of laser iridotomy. Am J Ophthalmol. 2005;140:1133-1136.
12. Radhakrishnan S, See J, Chew PT, et al. Illumination-induced changes in the angle configuration: an evaluation by anterior segment optical coherence tomography. IOVS. 2005;46:e-abstract 148.