OPTICAL COHERENCE TOMOGRAPHY
Overview
Since its commercial introduction in 1996, the clinical
applications of optical coherence tomography (OCT) have
expanded continuously. After nearly a decade of exploration
of OCT applications for use in the posterior segment,
the first dedicated anterior segment OCT (AS-OCT) system
debuted in 2005 (Visante OCT; Carl Zeiss
Meditec, Inc.). AS-OCT uses a longer wavelength (approximately
1,300 nm) than posterior segment OCT (approximately
840 nm), which permits deeper penetration and
can image from the cornea to the iris in one scan.
Preparation
Because AS-OCT is noncontact, the only preparation
required for using the device is to clean the chin and forehead
rests with alcohol prep pads and to enter patients'
information.
Assessment of the Cornea and Anterior Chamber
The basic type of scan is low-resolution sulcus-to-sulcus,
which provides clinicians with a comprehensive view of
the entire anterior chamber. I recommend performing not
only a single scan of the anterior segment (along one horizontal
axis) but also dual (two axes) and quad (four axes)
scans. High-resolution scans of the central cornea can also
be performed, especially for quantitative assessments such
as thickness measurements for LASIK flaps.
The OCT's signal is strongest when the scanning beam hits its target perpendicularly. Look for the center of the corneal surface, where the bright white lines appear horizontally together with a vertical white line (Figure 1). If these lines are not visible—even when the overall image is well aligned—adjust the scanner vertically by moving it slightly higher or lower as the scanning plane slices the target in an oblique fashion.
Assessment of the Angle
I recommend the high-resolution scan for optimal
visualization of the anterior segment angle. Ask the
patient to look sideways so that the nasal or temporal
side of the limbus is fully exposed to the scanner.
Because it is nearly impossible to install an effective
external fixation light on AS-OCT scanners, you must
work with patients to maximize the effect. The goal is
to visualize the interface between the sclera and ciliary
body as horizontally as possible (Figure 2). Because the
scleral spur defines one end of the trabecular meshwork,
the subjective assessment of the angle's opening
totally relies on its location. To maximize the visibility
of the scleral spur, ensure that the scanning beam hits
the limbus' surface as perpendicularly as possible.
Sometimes, the Visante's image preview screen shows a distorted cornea no matter how you adjust the scanning location and angle (Figure 3). This is because the device's image-visualization engine is optimized for scanning the cornea instead of the angle, but you can still subjectively assess narrow-angle conditions. If angle occludability is the only clinical question, I suggest ignoring the distortion, but I do not recommend making a quantitative assessment of distorted images.
Spectral-Domain OCT
The software of two commercial posterior segment
spectral-domain OCT (SD-OCT) devices (RTVue
[Optovue Inc.] and Cirrus HD-OCT [Carl Zeiss
Meditec, Inc.]) was recently upgraded to include imaging
of the anterior segment. These all-in-one OCT systems
image both the anterior and posterior segments.
Moreover, they bring the benefits of SD-OCT technology
to anterior segment imaging, namely faster scanning
(26 kHz) and a higher axial resolution (approximately
5 μm). High resolution makes it possible for
you to visualize not only the trabecular meshwork but
also Schlemm canal (Figure 4). Despite the differences
in OCT engines, you can directly apply the scanning
techniques described earlier to scanning the anterior
segment with SD-OCT. Because the wavelength of
SD-OCT's light source is optimized for scanning the
posterior segment, however, only the surface of the iris
can be visualized (Figure 4).
ULTRASOUND BIOMICROSCOPY
Overview
In 1991, Pavlin et al introduced ultrasound biomicroscopy
(UBM), a clinical application of highfrequency
ultrasound. UBM uses a higher range of
frequency (50-100 MHz) than conventional ultrasound
(5-20 MHz), achieving a higher axial resolution
(approximately 25 μm). Since its inception, UBM
dominated the field of anterior segment imaging until
2005, when the first commercial AS-OCT unit was
introduced.
Preparation
Most UBM devices use balloon-like disposable
probe covers, which contain distilled water that serves
as a coupling medium between a UBM probe and the
eye (Figure 5A). Unlike the conventional plastic eyecup
system, the probe cover system allows patients to be
scanned in a variety of postures—supine, sitting, and
even prone. For maximum maneuverability and probe
stability, choose the supine position.
Do not completely fill the probe cover with distilled water. The balloon should look a bit saggy so that you can minimize the pressure on the eye when applying the probe to the corneal surface (Figure 5B and 5C).
After instilling an anesthetic drop, place the little finger of the hand holding the probe on the lower eyelid, and place the thumb or middle finger of your other hand on the upper eyelid (Figure 6). With one smooth synchronized rotating motion of both hands, gently place the probe (covered by a water-filled balloon) on the cornea while applying gentle but firm tension on both the upper and lower eyelids. Finally, place the index finger of the hand holding the upper eyelid to the side of the probe above the balloon. This way, the hand that is holding the probe rests rigidly on the patient's cheek, and you can control the probe with a delicate but precise motion with support from the index finger of the other hand. This method also gives you a sense of the probe's location in threedimensional space, which is useful when you are performing UBM in a dark room (dark room provocative test for angle-closure glaucoma).
Probe Handling
The UBM probe is held freely in three-dimensional
space, which makes it challenging for you to master
6 degrees of freedom (Figure 7). I always ask beginners
to scan a plastic model eye to get a feel of each
degree of freedom one by one. Use forward-back and
left-right movements for coarse registration of the
region of interest (ROI) in the center of the viewing
screen. Adjust up-down for the vertical location of
the ROI within the viewing screen. Up-down is the key
movement to keep the ROI in focus. The focal plane is
perpendicular to the scanning beam. Think of this as
a horizontal band in the viewing screen. After registering
the ROI to the optimal scanning location, use the
other 3 degrees of freedom (pitch, yaw, and roll) to
fine-tune the angle to be scanned to the ROI.
Scanning for Glaucoma Assessment
It is essential to provide clear images of the anterior
chamber angle at any location in a uniform way so
that the physician can easily compare one image to
another. I recommend placing the scleral side on the
left of the screen and the corneal side on the right
(Figure 8). Ask the patient to look away from the ROI
so that the angle in question is fully exposed to the
probe. For example, if you want to scan the temporal
angle (9 o'clock for the right eye), ask the patient to
look to the nasal side (look to the left when the right
eye is scanned). Alternatively, to scan the 1-o'clock
region, ask the patient to look in the direction of
7 o'clock.
Scanning Tips
As with OCT, UBM also
generates the strongest signal
when the scanning beam hits
its target perpendicularly.
You therefore must carefully
choose an incident angle
depending on the structure
of focus. For narrow angle
cases, focus on the angle's
structure, especially on
the trabecular meshwork.
Try imaging the cornea as
horizontally as possible so that the scleral spur and
Schwalbe line are visible (Figure 8B). On the other
hand, for a case in which a tube is implanted behind the iris, image the iris as horizontally as possible so
that the tube can be seen in full length within the
scanning window (Figure 8C).
Scanning for Other Pathologies
UBM can also be used to assess tumors and cysts.
Most cases require size and extent measurements. To
demonstrate three-dimensional extent, image a longitudinal
(parallel to the limbus) slice through the lesion
in addition to a regular transverse (radial) slice. On
healthy eyes, a group of ciliary processes can be visualized
through the width of the scanning window on
successful longitudinal scans (Figure 8D).
